Terumo Advanced Perfusion System 1 Ultrasonic Air Sensor 1/4x1/16 UAS BLK, Catalog Number 5785 - Product Usage: intended for use with Terumo and Sarns brand air detection and heart-lung systems to detect gross air bubbles in the line during extracorporeal procedures.

Z-2960-2020

August 25, 2020

Class II

Product summary

Firm
Terumo Cardiovascular Systems Corporation
Event
Event 86338
Status
Terminated
Classification
Class II
Quantity
3
Official record key
device-enforcement:Z-2960-2020

Official wording

Reason: Internal testing at the supplier, revealed that a crimped wire was placed into an air bubble detector (ABD) sensor connector backwards, preventing the crimped wire from locking into the mating wire connector. This intermittent connection may result in false air bubble alarms.

Code information: UDI: 00886799000045; Serial Numbers: 20230 20263 20265

Distribution pattern: Worldwide distribution - US Nationwide distribution.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Internal testing at the supplier, revealed that a crimped wire was placed into an air bubble detector (ABD) sensor connector backwards, preventing the crimped wire from locking into the mating wire connector. This intermittent connection may result in false air bubble alarms.