Device product
Polaris 5.5 Spinal System and Polaris 5.5/Cypher Spinal System, Thoracolumbosacral pedicle screw system
Z-0570-2021
Product summary
- Event
- Event 86762
- Status
- Terminated
- Classification
- Class II
- Quantity
- 57,715 units
- Official record key
device-enforcement:Z-0570-2021
Official wording
Reason: Zimmer Biomet is conducting a medical device correction to update the Instructions for Use (IFU) and the Surgical Techniques for the Polaris Spinal System Translation Screw and Cypher MIS Screw System. Users should be aware that changes were made to the contraindications section indicating that the Translation Screws should not be used with a direct current stimulation, as a high probability of construct failure exists when using Translation Screws containing CoCr with direct current simulation system.
Code information: There have been 756 item numbers cleared in the US that are included in the recall scope, as listed on the first tab of the attached ZFA 2020-00333 Affected Product List. To date, there have only been products manufactured for a portion of the affected item numbers. In addition, there are three kit part numbers affected as well. The lot specific information is included on the second tab of the attached ZFA 2020- 0333 Affected Product List.
Distribution pattern: Domestic distribution nationwide. As direct stimulation devices are only available in the US, this contraindication does not apply to international customers.
Derived failure modes
-
Unknown
Zimmer Biomet is conducting a medical device correction to update the Instructions for Use (IFU) and the Surgical Techniques for the Polaris Spinal System Translation Screw and Cypher MIS Screw System. Users should be aware that changes were made to the contraindications section indicating that the Translation Screws should not be used with a direct current stimulation, as a high probability of construct failure exists when using Translation Screws containing CoCr with direct current simulation system.