Recall Observatory FDA recall evidence

Device product

MicroScan WalkAway-40 plus Reconditioned Instrument

Z-1372-2021

September 04, 2019

Class II

Product summary

Firm
Beckman Coulter Inc.
Event
Event 87486
Status
Terminated
Classification
Class II
Quantity
6 systems
Official record key
device-enforcement:Z-1372-2021

Official wording

Reason: Instruments were manufactured with a shortened 30mL reagent draw straw which may result in reagents not dispensing and may lead to organism misidentification and/or Quality Control (QC) failure.

Code information: Catalog Number: B1018-340R UDI Code: 15099590684112 Serial Numbers: 34020035 34020224 34030054 3407179 3407459 3407532

Distribution pattern: U.S.: CO, FL, IL, IN, KY, ME, MI, MS, PA, Puerto Rico, TN, TX, UT, VA and WA O.U.S.: India, Indonesia, Italy, Japan, Philippines, Poland, Russia, South Korea, Taiwan and Yemen.

Derived failure modes

  • Manufacturing or process control

    reason.manufacturing_process_control · v1.0.0

    Quality Control (QC) failure