Skip to content
Recall Observatory FDA recall evidence

Device product

NovaPACS Diagnostic Viewer. NovaPACS is a picture archiving and communication system software that retrieves, distributes, and display images and data from different modalities including CT, MR, US, CR/DX, NM, PT, and XA.

Z-2956-2020

September 10, 2019

Class II

Product summary

Firm
Novarad Corporation
Event
Event 85347
Status
Terminated
Classification
Class II
Quantity
197 units
Official record key
device-enforcement:Z-2956-2020

Official wording

Reason: The firm received a report of an atypical dataset being generated. When using the cross-localization feature and images from a modality that generates asymmetrical images, the cross-localization reference line may be inaccurately placed on any of the corresponding images that are open.

Code information: NovaPACS versions 8.6.8, 8.6.9, 8.6.11, 8.6.12, 8.6.14, 8.7.4, 8.7.6, and 8.7.8

Distribution pattern: Worldwide distribution - U.S. Nationwide distribution including in the states of AK, AZ, CA, CO, DC, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The countries of Canada, Colombia, Dominican Republic, Indonesia, Jordan, Mexico, Peru, Phillipines, and United Kingdom.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0