Device product

Inquiry Steerable Diagnostic Catheter, REF: IBI-81104 - Product Usage: INDICATIONS FOR USE The Inquiry fixed curve and steerable electrophysiology catheters are used for electrogram recording and cardiac stimulation during diagnostic electrophysiology studies. The catheters are commonly placed at the high right atrium, right ventricular apex, and HIS bundle.

Z-3011-2020

August 24, 2020

Class II

Product summary

Firm
Irvine Biomedical Inc, a St. Jude Medical Co.
Event
Event 86328
Status
Terminated
Classification
Class II
Quantity
15 units
Official record key
device-enforcement:Z-3011-2020

Official wording

Reason: Steerable diagnostic catheter packages labeled Model IBI-81104, batch 7397352 incorrectly contained Model IBI-87002, batch 7397307. Packages labeled Model IBI-87002, batch 7397307 incorrectly contained Model IBI-81104, batch 7397352.If physicians fail to observe the catheter difference, the intracardiac electrograms will not be representative of the desired signals, could result in procedure delay

Code information: GTIN: 05414734302975, Lot: 7397352

Distribution pattern: Worldwide distribution - U.S. Nationwide distribution including the states of MO, IN, WI, TX, AZ, MI, VA, KY and the country of Japan.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Steerable diagnostic catheter packages labeled Model IBI-81104, batch 7397352 incorrectly contained Model IBI-87002, batch 7397307. Packages labeled Model IBI-87002, batch 7397307 incorrectly contained Model IBI-81104, batch 7397352.If physicians fail to observe the catheter difference, the intracardiac electrograms will not be representative of the desired signals, could result in procedure delay