Pump Tube Set for MMS Libra and Solar Systems - Product Usage: Intended for use during water perfusion urodynamic procedures to study the function of the bladder and urethra during filling and voiding, by monitoring pressure and pressure changes.

Z-2535-2020

April 22, 2020

Class II

Product summary

Firm
LABORIE MEDICAL TECHNOLOGIES, CANADA ULC
Event
Event 85655
Status
Terminated
Classification
Class II
Quantity
480 units
Official record key
device-enforcement:Z-2535-2020

Official wording

Reason: This field correction is being initiated to address potentially damaged seals in the packaging of some the Laborie Pump Tube Sets. Finished goods inventory was detected with damage to their seals in the form of breaches or folds. Based on our investigation the issue appears to be limited to a single lot of part number 902101226 pump tubing sets.

Code information: Model Number: REF 902101226, LOT 15883/1

Distribution pattern: US Nationwide distribution including in the states of CO, FL, IN, MI, NJ, NY, OH, OR, PA and TX.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    This field correction is being initiated to address potentially damaged seals in the packaging of some the Laborie Pump Tube Sets. Finished goods inventory was detected with damage to their seals in the form of breaches or folds. Based on our investigation the issue appears to be limited to a single lot of part number 902101226 pump tubing sets.