Device product

CAS-930A Flat Panel Monitor (FPD) used in conjunction with the Canon Interventional Angiography System, Model numbers: Infinix-8000C and Infinix-8000H - Product Usage: This device is a digital radiography/fluoroscopy system used in a diagnostic and interventional angiography configuration. The system is indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen and lower extremities.

Z-0532-2021

October 08, 2020

Class II

Product summary

Firm
Canon Medical System, USA, INC.
Event
Event 86650
Status
Terminated
Classification
Class II
Quantity
14 units
Official record key
device-enforcement:Z-0532-2021

Official wording

Reason: There is a potential that the cable of the Flat panel detector may break during a procedure which may cause the x-ray device to stop before the examination is completed. This could result in a delay with procedure and/or potential damage to the patients blood vessels due to removal without visibility.

Code information: Affected Serial Numbers: 1 AFA1662001, 2 AFB1872005, 3 AFB1832003, 4 AFB1872004, 5 AFB1892006, 6 BFA18Y7001, 7 AEA17X2001, 8 AFA1692002, 9 AEB1812002, 10 BFA1932001, 11 BFB1952003, 12 BFB1942002, 13 BFB1972004, 14 BFB1972005, 15 AFB1892007, 16 BEC2032001, 17 BFC2062007, 18 BFC2062008, 19 BHA1992001, 20 BFC2012006

Distribution pattern: US Nationwide distribution including in the states of LA, CA, CO, WV, GA, NH, FL, PA, WI, MN, ID.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    There is a potential that the cable of the Flat panel detector may break during a procedure which may cause the x-ray device to stop before the examination is completed. This could result in a delay with procedure and/or potential damage to the patients blood vessels due to removal without visibility.