Device product

Normand Remisol Advance Data Manager is a software device that interfaces between laboratory information systems (LIS) and laboratory instruments.

Z-1831-2021

March 22, 2021

Class II

Product summary

Firm
Normand-Info S.A.S.U.
Event
Event 87750
Status
Terminated
Classification
Class II
Quantity
8,402 units with the affected software
Official record key
device-enforcement:Z-1831-2021

Official wording

Reason: Instructions For Use for software versions 1.5 to 1.9 did not provide sufficient instructions regarding the proper programming and testing of rules. Therefore, rules written by users that were not validated or incorrectly validated may lead to erroneous results.

Code information: Catalog Numbers: B08688; B22517; B35058; B74811; B92487 UDI: None Lot Numbers: All Lots

Distribution pattern: Worldwide distribution - US Nationwide distribution in the states of Alabama, Colorado, Connecticut, Delaware, District of Columbia, Florida, Georgia, Hawaii, Idaho, Indiana, Iowa, Kansas, Louisiana, Maine, Maryland, Massachusetts, Michigan, Mississippi, Missouri, Montana, Nebraska, Nevada, New Hampshire, New Jersey, New York, Pennsylvania, Rhode Island, South Carolina, Texas, West Virginia and Wyoming. The countries of Australia, Austria, Belgium, Brunei, Canada, China, Colombia, Congo, Czech Republic, Denmark, Ethiopia, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Lithuania, Malaysia, Mayotte, Morocco, Netherlands, Netherlands Antillean, New Zealand, Norway, Poland, Portugal, Qatar, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Uganda, UAE, UK, and Zimbabwe.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Instructions For Use for software versions 1.5 to 1.9 did not provide sufficient instructions regarding the proper programming and testing of rules. Therefore, rules written by users that were not validated or incorrectly validated may lead to erroneous results.