Device product
Alinity m System, Part Number: 08N53-002
Z-1602-2024
Product summary
- Event
- Event 94251
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 1,012 units
- Official record key
device-enforcement:Z-1602-2024
Official wording
Reason: The Alinity m System camera detects whether tubes are "capped." If the camera detects a cap and the retention bar is not present on the sample rack, the system will generate the message 'Error - 6013: Has cap, no retention bar." Abbott has identified an issue concerning the use of certain third-party Universal Transport Medium (UTM) tubes on the Alinity m System. This issue has been observed with Copan UTM-RT 3 mL Without Beads Transport & Preservation Medium for Viral Molecular Diagnostics Testing, Part Number (PN) 3C047N. However, it is possible that other third party manufactured tubes could experience the same issue.
Code information: Part Number: 08N53-002; UD/DI: 00884999048034. Not Serial Number specific
Distribution pattern: Worldwide distribution - US Nationwide and the countries ARGENTINA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BELGIUM, BOSNIA AND HERZEGOVINA, BRAZIL, BULGARIA, CAMBODIA, CAMEROON, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, CROATIA, CZECH REPUBLIC, EGYPT, EL SALVADOR, ESTONIA, ETHIOPIA, FINLAND, FRANCE, GERMANY, GREECE, HONG KONG, INDIA, IRELAND, ISRAEL, ITALY, JAPAN, KAZAKHSTAN, KENYA, LATVIA, LUXEMBOURG, MALAWI, MALAYSIA, MEXICO, MONTENEGRO, MOROCCO, MOZAMBIQUE, NETHERLANDS, NEW ZEALAND, NIGERIA, NORWAY, PAKISTAN, PANAMA, PARAGUAY, POLAND, PORTUGAL, ROMANIA, RUSSIA (CIS), SAUDI ARABIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, TANZANIA, THAILAND, TURKEY, UGANDA, UNITED KINGDOM, VIETNAM, ZIMBABWE.
Derived failure modes
-
Unknown
The Alinity m System camera detects whether tubes are "capped." If the camera detects a cap and the retention bar is not present on the sample rack, the system will generate the message 'Error - 6013: Has cap, no retention bar." Abbott has identified an issue concerning the use of certain third-party Universal Transport Medium (UTM) tubes on the Alinity m System. This issue has been observed with Copan UTM-RT 3 mL Without Beads Transport & Preservation Medium for Viral Molecular Diagnostics Testing, Part Number (PN) 3C047N. However, it is possible that other third party manufactured tubes could experience the same issue.