Device product
QUIDEL InflammaDry MMP-9 Test, REF: PRS-ID-20-U, 20 Tests
Z-1636-2024
Product summary
- Event
- Event 94219
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 104,300 units
- Official record key
device-enforcement:Z-1636-2024
Official wording
Reason: There are two windows on the test device, the control line window and the test results window. The line for the test results window were misaligned. In instances of positive test results, the positivite test line may not be visible/ hidden under the cassette of the test device thus potentially leading to a false negative result.
Code information: UDI: 30014613336754/Lot # 162621 162695 162498 162718 162749 163025
Distribution pattern: US: AK AL AR AZ CA CO CT DC FL GA HI IA ID IL IN KS KY LA MA MD MI MN MO MT NC ND NE NH NJ NV NY OH OK OR PA SC SD TN TX UT VA WA WI WV OUS: United Arab Emirates, Belgium, Canada, Chile, China, Germany, Spain, United Kingdom, Ireland, Italy, South Korea, Lithuania, Qatar
Derived failure modes
-
Unknown
There are two windows on the test device, the control line window and the test results window. The line for the test results window were misaligned. In instances of positive test results, the positivite test line may not be visible/ hidden under the cassette of the test device thus potentially leading to a false negative result.