Device product

Proteus235- Proton Therapy System: to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation

Z-1205-2022

April 20, 2022

Class II

Product summary

Firm
Ion Beam Applications S.A.
Event
Event 90213
Status
Ongoing
Classification
Class II
Quantity
4 units
Official record key
device-enforcement:Z-1205-2022

Official wording

Reason: First layer of Uniform Scanning treatment fields is sometimes irradiated with incorrect Scanning Magnet setpoints resulting in the field size may be smaller than expected and the irradiated area may receive more dose than prescribed

Code information: UDI-DI: (01)05404013801138 Serial Numbers: PAT108 (US), PAT112 (US), SAT116 (US). PAT107 (EU).

Distribution pattern: Worldwide distribution - US Nationwide distribution in the states of IL, VA, WA and the country of Germany.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    First layer of Uniform Scanning treatment fields is sometimes irradiated with incorrect Scanning Magnet setpoints resulting in the field size may be smaller than expected and the irradiated area may receive more dose than prescribed