Skip to content
Recall Observatory FDA recall evidence

Device product

Version: 1.2.0 (1.2.0+85ela410) Caption Health, Caption AI, Ultrasound Imaging System Application Software (GMDN: 40873), UDI: (01)00860003586303(10)1.2.0)

Z-2111-2021

May 19, 2021

Class II

Product summary

Firm
Caption Health
Event
Event 88101
Status
Terminated
Classification
Class II
Quantity
15 systems
Official record key
device-enforcement:Z-2111-2021

Official wording

Reason: The firm is aware of a issue with ultrasound software that results in unintended video frames being included. This could lead to an incorrect automated ejection fraction (AutoEF) and image quality score (IQS) estimates.

Code information: All product containing software version 1.2.0

Distribution pattern: US : AZ, DE, NY, IL, CT OUS: None

Derived failure modes

  • Device software or design

    reason.device_software_design · v1.0.0

    issue with ultrasound software