Skip to content
Recall Observatory FDA recall evidence

Device product

Naturalyte Liquid Acid concentrate drums, Product Code 13-2251-0 - Product Usage: used in conjunction with Fresenius Renal Technologies Sodium Bicarbonate Concentrate in a compatible 45X dilution three-stream artificial kidney (hemodialysis) machine.

Z-0839-2021

November 23, 2020

Class II

Product summary

Firm
Fresenius Medical Care Holdings, Inc.
Event
Event 86939
Status
Ongoing
Classification
Class II
Quantity
224 drums
Official record key
device-enforcement:Z-0839-2021

Official wording

Reason: The deviation in conductivity for sodium level exceeds 0.6mS/cm from the theoretical conductivity (TCD) or approximately 6mEq/L of sodium from prescriptions that was previously communicated. Calcium concentration is below manufacturing specification.

Code information: Lot Numbers: 20LXAC025 20LXAC081 20LXAC082 20LXAC083 Additional Affected lots as of 3/17/21: 20LXAC029 20NXAC021 20LXAC026 20LXAC065 20LXAC028 20NXAC029 20LXAC071 20LXAC103 20LXAC027 20LXAC070 20LXAC067 20NXAC019 20LXAC082 20LXAC025 20LXAC081 20LXAC083

Distribution pattern: US Nationwide distribution.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The deviation in conductivity for sodium level exceeds 0.6mS/cm from the theoretical conductivity (TCD) or approximately 6mEq/L of sodium from prescriptions that was previously communicated. Calcium concentration is below manufacturing specification.