Skip to content
Recall Observatory FDA recall evidence

Device product

uPath software 1.1 and uPath software 1.1 IVD-INSW, Material numbers 08991871001 and 09077332001

Z-0877-2021

November 23, 2020

Class II

Product summary

Firm
Roche Molecular Systems, Inc.
Event
Event 86996
Status
Terminated
Classification
Class II
Quantity
27
Official record key
device-enforcement:Z-0877-2021

Official wording

Reason: When a user creates a measurement tool annotation in the uPath Enterprise software version 1.1, the measurement value is incorrectly calculated when the measurement is viewed in "Split View" viewing mode and the slides within the case are scanned at different magnifications.

Code information: version 1.1

Distribution pattern: Domestic distribution to Maryland and Michigan. International distribution to France, Greece, Italy, Poland, Turkey, Brazil, India, and Singapore.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    When a user creates a measurement tool annotation in the uPath Enterprise software version 1.1, the measurement value is incorrectly calculated when the measurement is viewed in "Split View" viewing mode and the slides within the case are scanned at different magnifications.