Device product
11278VSA 11278VSAK Flexible Video-Uretero-Renoscope (FLEX-XC) Z23117US-BA (10/2020); 11278VSE 11278VSEK Flexible Video-Uretero-Renoscope (FLEX-XC) Z23117US-BA (10/2020); 11278VSU 11278VSUK Flexible Video-Uretero-Renoscope (FLEX-XC) Z18447US-BC (03/2018); 11278VSUA 11278VSUAK Flexible Video-Uretero-Renoscope (FLEX-XC) Z23117US-BA (10/2020); 11278VSUE 11278VSUEK Flexible Video-Uretero-Renoscope (FLEX-XC) Z23117US-BA (10/2020)
Z-1070-2022
Product summary
- Event
- Event 89880
- Status
- Ongoing
- Classification
- Class II
- Quantity
- All Serial Numbers manufactured/distributed since January 2018.
- Official record key
device-enforcement:Z-1070-2022
Official wording
Reason: Failure to achieve the expected six-log reduction in microorganisms following the disinfection process.
Code information: 11278VSA 11278VSAK Flexible Video-Uretero-Renoscope (FLEX-XC) Z23117US-BA (10/2020) UDI Code: 4048551402597; 11278VSE 11278VSEK Flexible Video-Uretero-Renoscope (FLEX-XC) Z23117US-BA (10/2020) UDI Code: 4048551416976; 11278VSU 11278VSUK Flexible Video-Uretero-Renoscope (FLEX-XC) Z18447US-BC (03/2018) UDI Code: 4048551344385; 11278VSUA 11278VSUAK Flexible Video-Uretero-Renoscope (FLEX-XC) Z23117US-BA (10/2020) UDI Code: 4048551402603; 11278VSUE 11278VSUEK Flexible Video-Uretero-Renoscope (FLEX-XC) Z23117US-BA (10/2020) UDI Code: 4048551416969; All Serial Numbers manufactured/distributed since January 2018.
Distribution pattern: U.S.: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA,WI, WV, and WY. O.U.S.: Mexico, Lebanon, and South Korea
Derived failure modes
-
Unknown
Failure to achieve the expected six-log reduction in microorganisms following the disinfection process.