ASC PneumoLiner device, Part No. WA90500US (US market only) The Pneumoliner device is intended for use as a multiple instrument port and tissue containment system during minimally invasive gynecologic laparoscopic surgery to enable the isolation and containment of tissue, considered benign, resected during single-port or multi- site laparoscopic surgery during power morcellation and removal.

Z-0176-2022

September 21, 2021

Class II

Product summary

Firm
Olympus Corporation of the Americas
Event
Event 88750
Status
Terminated
Classification
Class II
Quantity
350 pieces/70 boxes
Official record key
device-enforcement:Z-0176-2022

Official wording

Reason: Due to manufacturing error, the Pneumoliner Bag Distal Tab that exits the Introducer shaft is in the wrong orientation. This results in the user deploying the bag upside down, which will make tissue encapsulation and bag closure more difficult, introducing the risk of trapping the small bowel/viscera in the bag at closure resulting in patient injury.

Code information: Lots 647572 and 667060

Distribution pattern: US Nationwide distribution to states of: AL, CA, CT, FL, IL, MD, MI, MN, MO, NE, NY, OH, OK, OR, PA, SC, UT, VA, WA, WI, and WV ; and International (foreign) distribution to: Europe.

Derived failure modes

  • Manufacturing or process control

    reason.manufacturing_process_control · v1.0.0

    manufacturing error