Device product
ExacTrac Dynamic (Medical Charged-Particle Radiation Therapy System).
Z-1185-2021
Product summary
- Firm
- Brainlab AG
- Event
- Event 87395
- Status
- Terminated
- Classification
- Class II
- Quantity
- 73 ExacTrac Dynamic systems (versions 1.0.0, 1.0.1, 1.0.2) have been manufactured and distributed
- Official record key
device-enforcement:Z-1185-2021
Official wording
Reason: In case of a failed automatic marker detection, a software error causes parts of the display to incorrectly behave as if the current patient position is within predefined tolerances and may allow the user to proceed to treatment despite potentially exceeding shift values.
Code information: Model/catalogue numbers: 20910-06 ETD Implanted lic. UDI: The GTIN for ExacTrac Dynamic version 1.0 is 04056481142315. GMDN code: 40887. Software revisions of affected ExacTrac Dynamic software: versions 1.0.0, 1.0.1, 1.0.2.
Distribution pattern: USA: (CA, NJ, AZ, AR, NY, IN, OR, NC), Australia, Austria, Belgium, Denmark, France, Germany, Hong Kong, Italy, Slovakia, Spain, Thailand, Turkey, United Kingdom, United Arab Emirates.
Derived failure modes
-
Device software or design
software error