inLight Medical LED Pads used with Polychromatic light technology system
Z-1943-2021
Product summary
- Event
- Event 85106
- Status
- Terminated
- Classification
- Class III
- Quantity
- 926
- Official record key
device-enforcement:Z-1943-2021
Official wording
Reason: LED pads with cord connectors that have v-shaped notches may not have been assembled properly. This issue could lead to a power interruption, causing the controller and power supply to shut down.
Code information: All serial numbers with cord connectors that have v-shaped notches
Distribution pattern: Worldwide distribution - US Nationwide distribution in the states of CA, TX, NC, ID, CO, VA, FL, VT, AK, DC, MA, IL, AZ, MN, IA, TN, WI, WA, OK, MI, OH, GA, NY, IN, PA, NV, NJ, AR, CO , LA, MT, OR, SD, ND, MD, UT, MO, KY, NM, KS, AS, NE, WY, PR, MS and the country of Canada.
Derived failure modes
-
Unknown
LED pads with cord connectors that have v-shaped notches may not have been assembled properly. This issue could lead to a power interruption, causing the controller and power supply to shut down.