Device product
HeartSpan Transseptal Needles Catalog No. FND-019-01, Lot No. E1913644 consisting of an outer needle cannula and an inner stylet. The needle is comprised of flexible thin walled tubing with an ergonomic hub and stopcock attached to the proximal end. The stylet consists of a solid wire that when inserted in the needle protrudes beyond the distal tip of the cannula.
Z-1262-2021
Product summary
- Event
- Event 87306
- Status
- Terminated
- Classification
- Class II
- Quantity
- 587 Units
- Official record key
device-enforcement:Z-1262-2021
Official wording
Reason: The labeled needle tip curvature of specific lots of transseptal needles may not match the actual needle tip curvature. In the event that the incorrect needle is unknowingly used, an unintended anatomy puncture may occur, which is likely to require medical intervention.
Code information: Catalog Number FND-019-01, Lot Number E1913644; UDI: (01)00884450489345
Distribution pattern: U.S. Consignees: VA, NE, Washington DC, MI, NY, KY, CA, TX, MD, AL, GA, UT, LA, CA, CT, PA, WA, MT and NJ. No government consignees. Also distributed OUS.
Derived failure modes
-
Unknown
The labeled needle tip curvature of specific lots of transseptal needles may not match the actual needle tip curvature. In the event that the incorrect needle is unknowingly used, an unintended anatomy puncture may occur, which is likely to require medical intervention.