Access SARS-CoV-2 lgG II Calibrator, REF C69058, IVD, CE UDI: (01)15099590742751

Z-1906-2021

May 07, 2021

Class II

Product summary

Firm
Beckman Coulter Inc.
Event
Event 87960
Status
Terminated
Classification
Class II
Quantity
3,843 units
Official record key
device-enforcement:Z-1906-2021

Official wording

Reason: SARS-CoV-2 lgG II numerical results may potentially be multiplied by a factor of 1000 on systems running with assay protocol file (APF) and access assay file (AAF) versions resulting in falsely elevated numerical values.

Code information: All Lots

Distribution pattern: US: Alabama, Alaska, Arizona, Arkansas, California, Florida, Georgia, Illinois, Iowa, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Mississippi, Missouri, New Hampshire, New Jersey, New York, North Dakota, Ohio, Oklahoma, Pennsylvania, Puerto Rico, Tennessee, Texas, Virginia, West Virginia, Wisconsin, Wyoming, OUS: Andorra, Austria, Croatia, Czech Republic, France, Germany, Gibraltar, Greece, Hong Kong, Hungary, Israel, Italy, Japan, Jordan, Poland, Portugal, San Marino, Slovakia, Slovenia, Spain, United Arab Emirates

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    SARS-CoV-2 lgG II numerical results may potentially be multiplied by a factor of 1000 on systems running with assay protocol file (APF) and access assay file (AAF) versions resulting in falsely elevated numerical values.