Device product
APEX Compounding System Control Panel Module, REF: 601224, Product Code AX1000 - Product Usage: intended for intravascular administration to a patient.
Z-0717-2021
Product summary
- Event
- Event 86803
- Status
- Terminated
- Classification
- Class II
- Quantity
- 248 units
- Official record key
device-enforcement:Z-0717-2021
Official wording
Reason: There is the potential for the compounding system to not immediately interrupt compounding and alert user to the presence of an air bubble exceeding 3% of the ordered volume for an ingredient. This could result in a a compounded drug that does not meet the specified accuracy range for an individual ingredient.
Code information: All APEX Compounding System with Software Version 1.0.5.57 or earlier
Distribution pattern: US Nationwide distribution including in the states of AL, AZ, CA, CO, CT, FL, GA, IA, IL, LA, MA, MD, MI, MN, MO, NC, NH, NJ, NV, NY, OH, OK, PA, TN,TX, WA.
Derived failure modes
-
Unknown
There is the potential for the compounding system to not immediately interrupt compounding and alert user to the presence of an air bubble exceeding 3% of the ordered volume for an ingredient. This could result in a a compounded drug that does not meet the specified accuracy range for an individual ingredient.