Skip to content
Recall Observatory FDA recall evidence

Device product

Permobil TiLite, Models: Aero Z, and ZRA

Z-0264-2022

September 24, 2021

Class II

Product summary

Firm
TiSport, Llc
Event
Event 88718
Status
Ongoing
Classification
Class II
Quantity
5373
Official record key
device-enforcement:Z-0264-2022

Official wording

Reason: Some affected wheelchairs, with transit tie downs, that are used as a passenger seat in a motor vehicle may not be configured in the right orientation to perform as intended if involved in a motor vehicle accident. Some affected wheelchairs can no longer be used as a passenger seat until transit tie downs are replaced with redesigned transit tie downs, verified, and released to the market.

Code information: Manufactured between January 19, 2017 and June 14, 2021 with the serial number range from M2005054 to 1231112011

Distribution pattern: US Nationwide Distribution: MD, PA, VA, IN, NY, MO, AR, NE, NJ, WA, NM, IL, CA, OH, KY, LA, OK, WI, TX, NC, SD, MN, MA, OR; OUS (Foreign): AUSTRALIA, CANADA, FRANCE, GERMANY, IRELAND, ITALY, UNITED KINGDOM, URUGUAY, NEW ZEALAND, IRELAND, SWEDEN

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Some affected wheelchairs, with transit tie downs, that are used as a passenger seat in a motor vehicle may not be configured in the right orientation to perform as intended if involved in a motor vehicle accident. Some affected wheelchairs can no longer be used as a passenger seat until transit tie downs are replaced with redesigned transit tie downs, verified, and released to the market.