Skip to content
Recall Observatory FDA recall evidence

Device product

Pilot COVID-19 At-Home Test, Reference Number 9901-NCOV-10G, Catalog Number 99COV200L-EN02, UPC Code: 887473000207 (EAN-13: 08800111707728)

Z-1594-2023

March 31, 2023

Class I

Product summary

Firm
SD Biosensor, Inc.
Event
Event 92137
Status
Ongoing
Classification
Class I
Quantity
2,712,767 kits
Official record key
device-enforcement:Z-1594-2023

Official wording

Reason: The liquid buffer component in the affected test kit lots was determined to have bacterial contamination. User may notice a slight odor when the buffer is brought close to the nose.

Code information: Lot Numbers: 53K41T5T1 53K41X1T1 53K41X2T1 53K41X3T1 53K4211T1 53K4212T1 53K4213T1 53K4221T1 53K4222T1 53K4223T1 53K4224T1 53K4225T1 53K4231T1 53K4232T1 53K4233T1 53K4261T1 53K4262T1 53K4271T1 53K4272T1 53K4273T1 53K4274T1 53K4291T1 53K4292T1 53K42A1T1 53K42A2T1 53K42A3T1 53K42E1T1 53K42G1T1 53K42G2T1 53K42H1T1 53K42H2T1 53K42L1T1 53K42L2T1 53K4361AC 53K4362AC 53K4392AC 53K38N3T1 53K38N2T1 53K38N1T1 53K38N4T1 53K38N5T1 53K38P1T1 53K38P2T1 53K38P3T1

Distribution pattern: US Nationwide.

Derived failure modes

  • Microbial contamination

    reason.microbial_contamination · v1.0.0

    bacterial contamination