Device product
Torque Limiting Adaptor, AO Connect, 4 N-m, part #027-203, reusable, a component of the Bodycad Fine Osteotomy Implant and Disposable Instrumentation Tray, part #025-004.
Z-0244-2022
Product summary
- Event
- Event 88878
- Status
- Terminated
- Classification
- Class II
- Quantity
- 10 in the U.S.
- Official record key
device-enforcement:Z-0244-2022
Official wording
Reason: There is a risk of breakage of the torque limiting adaptor if used counterclockwise and it must not be used with a power tool due to no test data validating this use.
Code information: Torque Limiting Adaptor - Lot number SB000430. Fine osteotomy tray - Lot numbers SB000251, SB000252, SB000253, SB000254, SB000259, SB000261, SB000263, SB000264, SB000272, and SB000273.
Distribution pattern: Distribution was made to CO, MI, and NY. Foreign distribution was made to Canada and the United Kingdom. There was no military/government distribution.
Derived failure modes
-
Unknown
There is a risk of breakage of the torque limiting adaptor if used counterclockwise and it must not be used with a power tool due to no test data validating this use.