Skip to content
Recall Observatory FDA recall evidence

Device product

Torque Limiting Adaptor, AO Connect, 4 N-m, part #027-203, reusable, a component of the Bodycad Fine Osteotomy Implant and Disposable Instrumentation Tray, part #025-004.

Z-0244-2022

July 02, 2021

Class II

Product summary

Firm
Laboratoires Bodycad, Inc.
Event
Event 88878
Status
Terminated
Classification
Class II
Quantity
10 in the U.S.
Official record key
device-enforcement:Z-0244-2022

Official wording

Reason: There is a risk of breakage of the torque limiting adaptor if used counterclockwise and it must not be used with a power tool due to no test data validating this use.

Code information: Torque Limiting Adaptor - Lot number SB000430. Fine osteotomy tray - Lot numbers SB000251, SB000252, SB000253, SB000254, SB000259, SB000261, SB000263, SB000264, SB000272, and SB000273.

Distribution pattern: Distribution was made to CO, MI, and NY. Foreign distribution was made to Canada and the United Kingdom. There was no military/government distribution.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    There is a risk of breakage of the torque limiting adaptor if used counterclockwise and it must not be used with a power tool due to no test data validating this use.