Device product
Mevion S250 and S250i Proton Therapy Systems hand pendant (P/N 10263 or P/N 37808)
Z-0827-2022
Product summary
- Event
- Event 89758
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 39 (36 US, 3 OUS)
- Official record key
device-enforcement:Z-0827-2022
Official wording
Reason: Malfunction of the hand pendant controlling movement of the therapy couch and nozzle may result in unintentional motion. Patient injury is possible if they come in contact with the nozzle or other objects in the treatment room.
Code information: All theryapy systems, all hand pendant serial numbers. Therapy system UDI 00864366000100 and UDI 00864366000124.
Distribution pattern: Worldwide distribution - US Nationwide distribution in the states of FL, MO, NJ, OH, OK, UT, DC and the country of Netherlands.
Derived failure modes
-
Unknown
Malfunction of the hand pendant controlling movement of the therapy couch and nozzle may result in unintentional motion. Patient injury is possible if they come in contact with the nozzle or other objects in the treatment room.