1. Sensis, material # 10764561; UDI: 04056869010137 2. Sensis Vibe Hemo, material # 11007641; UDI: 04056869010199 3. Sensis Vibe Combo, material # 11007642; UDI: 04056869010205 The Sensis recording system is a diagnostic and administrative tool supporting hemodynamic catheterizations and/or electrophysiology studies, for cardiac, interventional Radiology, and surgical studies.

Z-0499-2022

November 23, 2021

Class II

Product summary

Firm
Siemens Medical Solutions USA, Inc
Event
Event 89199
Status
Ongoing
Classification
Class II
Quantity
1047
Official record key
device-enforcement:Z-0499-2022

Official wording

Reason: Software error which affects Sensis Vibe Hemo, Sensis and Sensis Vibe Combo systems with software version VD12A, which could cause possible hazard to patients, operators, or other persons and equipment. Under certain sporadic circumstances, the CO (Cardiac Output) measurement using the Thermodilution method will temporarily no longer be possible.

Code information: 1. Sensis- 10764561; UDI: 04056869010137; Serial Numbers: 103493, 103432, 103466 2. Sensis Vibe Hemo, material # 11007641; UDI: 04056869010199; Serial Numbers: 103137, 103144, 103453, 103108, 103111, 103047, 103048, 103171, 103182, 103193, 103301, 103170, 103223,, 103225, 103247, 103313, 103315, 103196, 103198, 103166, 103173, 103177, 103184, 103186, 103199, 102326, 103139, 103161, 103167, 103168, 103172, 103178, 100263, 103424, 103319, 103320, 103321, 103323, 103325, 103328, 103330, 103333, 103334, 103335, 103344, 100216, 103337, 100416, 102008, 103413, 101083, 100208, 100209, 102138, 102145, 102147, 103160, 103314, 102199, 103079, 102221, 103123, 103124, 103131, 100163, 100306, 100309, 103300, 101099, 101100, 101101, 103265, 100422, 103019, 103331, 103362, 103393, 103402, 103419, 103252, 103261, 103267, 103269, 103270, 103276, 103277, 103280, 103281, 103022, 103381, 103194 103145, 103148, 103152, 100345, 102183, 103345, 103368, 103415, 103416, 103245, 103249, 103250, 103264, 103080, 103082, 103086, 103088, 103089, 103090, 100231, 103431, 103306, 103341, 103114, 100070, 100071, 100074, 102155, 102157, 103040, 103042, 103039, 102174, 102182, 103130, 103200, 103234, 103116, 103117, 103119, 103288, 103008, 103216, 103465, 103189, 103208, 102010, 102295, 102296, 102302, 100006, 100005, 103352, 103283, 103308, 103310, 103115, 103244, 103246, 103014, 103242, 102143, 103359, 103274, 103437 3. Sensis Vibe Combo, material # 11007642; UDI: 04056869010205; Serial Numbers: 123021, 123004, 123047, 123053, 122011 ***Updated 1/18/22*** SENSIS Vibe Hemo, Serial #103185 ***Updated 2/24/22*** SENSIS Vibe Hemo Serial #103290 103510 103451 103461 103444 103450 103448 103445 103442 103449 103439 103457 SENSIS Vibe Combo Serial #123055 123056 123058 The following serial numbers have been removed from the scope of this recall: SENSIS Vibe Hemo Serial #103301 103424 ***Updated 8/5/22*** The following serial numbers have been added to the recall: SENSIS Vibe Combo Serial numbers 103307, 103312

Distribution pattern: US Nationwide distribution in the states of AK, AL, AR, AZ, CA, FL, GU, HI, IA, IL, IN, KS, KY, MA, MI, MO, MS, NC, NE, NJ, NY, OH, OK,PA, PR, SD, TN, TX, WA, WV.

Derived failure modes

  • Device software or design

    reason.device_software_design · v1.0.0

    Software error