Device product
Randox Cholesterol- For the quantitative in vitro determination of Cholesterol in serum and plasma. Catalog Number: CH2O0
Z-1424-2022
Product summary
- Event
- Event 90522
- Status
- Terminated
- Classification
- Class II
- Quantity
- 15 kits
- Official record key
device-enforcement:Z-1424-2022
Official wording
Reason: Product fails to meet the performance claims quoted on the kit insert: Manual Procedure when calibrating using the standard provided in the kit, the change in absorbance is lower than expected generating internal quality control results high outside range; Cholesterol (CHOL) Instrument Specific Application using the recommended calibration material CAL2351, the product fails to meet its linearity performance claim quoted on the kit inserts. The linearity is reduced by up to -35% however Quality Control results will still fall within assigned ranges. This may lead to delayed results.
Code information: GTIN : 05055273201130 Lot Number: 586177 Expiry Date: 28th Jan 2024
Distribution pattern: Nationwide Distribution - MD, OR, PR, MT, WV
Derived failure modes
-
Unknown
Product fails to meet the performance claims quoted on the kit insert: Manual Procedure when calibrating using the standard provided in the kit, the change in absorbance is lower than expected generating internal quality control results high outside range; Cholesterol (CHOL) Instrument Specific Application using the recommended calibration material CAL2351, the product fails to meet its linearity performance claim quoted on the kit inserts. The linearity is reduced by up to -35% however Quality Control results will still fall within assigned ranges. This may lead to delayed results.