Skip to content
Recall Observatory FDA recall evidence

Device product

LYFO Disk

Z-0788-2022

July 17, 2015

Class III

Product summary

Firm
Microbiologics Inc
Event
Event 89513
Status
Completed
Classification
Class III
Quantity
3 units
Official record key
device-enforcement:Z-0788-2022

Official wording

Reason: LYFO Disk Catalog #0983L is to contain Candida tropicalis. However, lot number 983-40-3 under Catalog #0983L, labeled as Candida tropicalis, is actually Yersinia entercolitica. This labeling error may cause a user's quality control to fail, resulting in testing needing to be repeated and delayed test results.

Code information: Catalog Number 0983L, UDI: 10845357023930, Lot Number (Expiration Date): 983-40-3 (02/11/2016)

Distribution pattern: Distribution is nationwide with three consignees located in Colorado, Georgia, and Ohio.

Derived failure modes

  • Labeling or packaging

    reason.labeling_packaging · v1.0.0

    labeling error