Skip to content
Recall Observatory FDA recall evidence

Device product

DxU 850m Iris Urine Microscopy Analyzer Catalog Number C49513, in-vitro diagnostic device, urine analysis

Z-1597-2023

April 05, 2023

Class II

Product summary

Firm
Beckman Coulter, Inc.
Event
Event 92056
Status
Ongoing
Classification
Class II
Quantity
107 units
Official record key
device-enforcement:Z-1597-2023

Official wording

Reason: Beckman Coulter has become aware of an intermittent issue in which the optional flag "Previous Sample Had Sperm" was enabled but not displayed so that a carryover event could have been investigated.

Code information: UDI/DI 15099590748647, All analyzers with Windows 10, Software Versions 8.5, 8.5.1 and 8.6

Distribution pattern: Worldwide Distribution

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Beckman Coulter has become aware of an intermittent issue in which the optional flag "Previous Sample Had Sperm" was enabled but not displayed so that a carryover event could have been investigated.