Device product
PushTracker E2 component utilized with the SmartDrive MX2+ Power Assist Device (Part Numbers: MX2-32K, MX2-32P, MX2-32PK, MX2-33D, MX2-33P, MX2-33PK, MX2-33S, MX2-150, MX2-167, MX2-32D); a wearable watch that communicates with a SmartDrive Power Assist Device via the SmartDrive MX2+ Application
Z-2004-2023
Product summary
- Firm
- Permobil
- Event
- Event 92250
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 18,905 units
- Official record key
device-enforcement:Z-2004-2023
Official wording
Reason: When multiple processes are running on the watch's Central Processing Unit (CPU), the application may fail unexpectedly. If this happens, the motor on the power assist device continues to run and the user may not be able to stop the device using tap gestures.
Code information: All units with software versions 1.0.00 to 1.0.04
Distribution pattern: Worldwide distribution.
Derived failure modes
-
Unknown
When multiple processes are running on the watch's Central Processing Unit (CPU), the application may fail unexpectedly. If this happens, the motor on the power assist device continues to run and the user may not be able to stop the device using tap gestures.