Skip to content
Recall Observatory FDA recall evidence

Device product

Evita V500 Ventilator with Installed CO2 Measurement Option-Intended for the ventilation of adult,pediatric and neonatal patients. Catalog Number: 8416400

Z-0108-2022

September 10, 2021

Class II

Product summary

Firm
Draeger Medical, Inc.
Event
Event 88635
Status
Ongoing
Classification
Class II
Quantity
6,797 units
Official record key
device-enforcement:Z-0108-2022

Official wording

Reason: Software 2.51 and Lower with Installed CO2 Measurement Option, may result in Restart of the Ventilator, brief cessation of ventilation can occur and loss of PEEP for approx. 8 seconds

Code information: Software 2.51 and Lower

Distribution pattern: US Nationwide

Derived failure modes

  • Device software or design

    reason.device_software_design · v1.0.0

    Software 2.51 and Lower with Installed CO2 Measurement Option, may result in