Skip to content
Recall Observatory FDA recall evidence

Device product

MEDLINE UNITE FOOT & ANKLE Screw, Non-Locking, Mini T8 (2.0x8mm), REF MMSN2008

Z-1045-2023

December 22, 2022

Class II

Product summary

Firm
MEDLINE INDUSTRIES, LP - Northfield
Event
Event 91491
Status
Ongoing
Classification
Class II
Quantity
420 units
Official record key
device-enforcement:Z-1045-2023

Official wording

Reason: The recall is due to observed intra-operative screw failure. When pre-drilled with ¿1.3mm drill bit for ¿2.0mm screws and ¿1.6mm drill bit for ¿2.4mm we have observed screw failure due to excessive torque.

Code information: GTIN 10193489120387, Lot Number 810304

Distribution pattern: US Nationwide distribution.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The recall is due to observed intra-operative screw failure. When pre-drilled with ¿1.3mm drill bit for ¿2.0mm screws and ¿1.6mm drill bit for ¿2.4mm we have observed screw failure due to excessive torque.