Device product
Leica Surgical Operating Microscopes, PROvido Part Numbers: a. 10448976, b. 10448950
Z-1851-2022
Product summary
- Event
- Event 90818
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 32 devices
- Official record key
device-enforcement:Z-1851-2022
Official wording
Reason: During internal testing, Leica Microsystems has become aware of a component change on the photodiodes inside the M530 Optics Carrier which will result in an inaccurate adjustment of the illumination limits by the software of the surgical microscope when "BrightCare Plus" with Luxmeter is used.
Code information: a. 10448976, Serial Numbers: 120322001, 300122001, 170921001, 301221001, 40122001, 140522001, 151221002, 301021002, 81221002, 100222002, 40222002, 220821001, 300122002, 110322001, 50322001, 180921001, 260821001, 90622001, 120422001, 100522001, 90522001, 210821001, 250122001, 301021001, 100222001; b. 10448950, Serial Numbers: 160222001, 90721001, 50422001, 50422002, 91221001, 160921001, 61221001
Distribution pattern: US Nationwide distribution.
Derived failure modes
-
Unknown
During internal testing, Leica Microsystems has become aware of a component change on the photodiodes inside the M530 Optics Carrier which will result in an inaccurate adjustment of the illumination limits by the software of the surgical microscope when "BrightCare Plus" with Luxmeter is used.