Skip to content
Recall Observatory FDA recall evidence

Device product

exoplan version 3.1 Rijeka Software -A medical software, intended to support the pre-operative planning of dental implants using the visualization of the implant placement within images of the patents anatomy.

Z-2633-2023

August 17, 2023

Class II

Product summary

Firm
Exocad GmbH
Event
Event 92921
Status
Ongoing
Classification
Class II
Quantity
96 licenses
Official record key
device-enforcement:Z-2633-2023

Official wording

Reason: A software library filtering error has been discovered which occurs when users are using the Step-by-Step Full Drill Protocol export functionality of exoplan 3.1 Rijeka. When a guided surgery treatment approach is selected, along with a Step-by-Step Full Drill Protocol export implant library, exoplan software does not filter out unsupported sleeve height positions for a particular sub-full drill protocol and instead shows ALL possible sleeve height positions for all sub-full drill protocols. The use of an improper sleeve height position in a surgical guide could result in patient injury

Code information: UDI-DI: (01)4260521365026(10)A03B01E8439 (US) (01)4260521365026(10)A03B01E8588 (SR1 US) (01)4260521365026(10)A03B01E8606 (SR1 US offline) Software xoplan 3.1 Rijeka

Distribution pattern: Worldwide distribution - US Nationwide and the country of Canada.

Derived failure modes

  • Device software or design

    reason.device_software_design · v1.0.0

    software library filtering error