Device product

Leica MICROSYSTEMS Proveo 8, Part Number 10449058, surgical microscope system

Z-0475-2023

November 21, 2022

Class II

Product summary

Firm
Leica Microsystems, Inc.
Event
Event 91200
Status
Ongoing
Classification
Class II
Quantity
43 units
Official record key
device-enforcement:Z-0475-2023

Official wording

Reason: There is a potential issue resulting in the Power Supply potentially becoming overheated due to internal components overstressing. Consequently, due to the overstressing of the internal components, one specific component can fail which will cause the device to unexpectedly shutdown.

Code information: Serial Numbers: 20522001, 20722001, 21221001, 40222001, 40722001, 70722001, 80522002, 110722001, 150422001, 180322001, 211021001, 260722001, 290422001, 300122001, 270522001. ***EXPANSION 09/08/2023, Serial Numbers: 271222001, 100423002, 31022001, 180623001, 71222001, 210523002, 270922002, 120623002, 100123001, 130623001, 290922002, 300922002, 210223001, 220223001, 251122001, 281122001, 20623002, 60922001, 200822001, 40523001, 40523002, 270922001, 61122002, 20423001, 300523002, 190822002, 300922001, 280922001.

Distribution pattern: US Nationwide

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    There is a potential issue resulting in the Power Supply potentially becoming overheated due to internal components overstressing. Consequently, due to the overstressing of the internal components, one specific component can fail which will cause the device to unexpectedly shutdown.