Device product
Leica MICROSYSTEMS Proveo 8, Part Number 10449058, surgical microscope system
Z-0475-2023
Product summary
- Event
- Event 91200
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 43 units
- Official record key
device-enforcement:Z-0475-2023
Official wording
Reason: There is a potential issue resulting in the Power Supply potentially becoming overheated due to internal components overstressing. Consequently, due to the overstressing of the internal components, one specific component can fail which will cause the device to unexpectedly shutdown.
Code information: Serial Numbers: 20522001, 20722001, 21221001, 40222001, 40722001, 70722001, 80522002, 110722001, 150422001, 180322001, 211021001, 260722001, 290422001, 300122001, 270522001. ***EXPANSION 09/08/2023, Serial Numbers: 271222001, 100423002, 31022001, 180623001, 71222001, 210523002, 270922002, 120623002, 100123001, 130623001, 290922002, 300922002, 210223001, 220223001, 251122001, 281122001, 20623002, 60922001, 200822001, 40523001, 40523002, 270922001, 61122002, 20423001, 300523002, 190822002, 300922001, 280922001.
Distribution pattern: US Nationwide
Derived failure modes
-
Unknown
There is a potential issue resulting in the Power Supply potentially becoming overheated due to internal components overstressing. Consequently, due to the overstressing of the internal components, one specific component can fail which will cause the device to unexpectedly shutdown.