Device product
LFS Flat screen arm system with surgical lamp for Models Aurora series, Aurora 2 series, Aurora 3 series, Aurora 4 series, Aurora Astro series, Stellar XL, and Stellar series, LFS Flat Screen Arm System Model Numbers LFSST2323, LFSLFSST29TV23, LFSST29TV23, LFSST23TV, LFSST23, LFSST2323TV, LFSLFSST2323, LDFSLFSLED7TV5, LDFSLFSLED7TV55, LFSLED55, LFSLFSST23, LFSST2923TV, LFSST29TV, LFSST29, LFSST19WC, LFSST2923, LFSLFSST-1, LFSST292323, LFSST1919WC, LFS ST2319WC
Z-0071-2024
Product summary
- Event
- Event 92999
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 256 units
- Official record key
device-enforcement:Z-0071-2024
Official wording
Reason: The stopper block on the LFS bracket could be broken or missing. Without this stopper block, the flat screen attached to the LFS flat screen system can rotate beyond the original angle limit and it could cause excessive force to be given to the spring shaft of the LFS bracket. Due to the excessive force given to the spring shaft, the spring shaft could break and shear off.
Code information: Serial Numbers: 06087323 - 06087352, 06087353 - 06087354, 06081292 - 06081293, 06087448 - 06087449, 06087450 - 06087469, 06097595 - 06097596, 06107716 - 06107718, 06107719 - 06107728, 06107730, 06107731 - 06107732, 06111348 - 06111349, 06117855 - 06117856, 06117857 - 06117866, 06117868 - 06117868, 06111350 - 06111351, 06117949 - 06117949, 06117954 - 06117973, 06118031 - 06118032, 06128116 - 06128131, 06128172 - 06128176, 06128177 - 06128186, 06128187 - 06128188, 07010004, 07010002, 07010005 - 07010006, 07028436 - 07028441, 07030080 - 07030081, 07048491 - 07048495, 07058506 - 07058507, 07058508 - 07058512, 07068602 - 07068604, 07068605 - 07068607, 07068611 - 07068622, 07068609 - 07068610, 07068667 - 07068670, 07061479 - 07061484, 07078679 - 07078680, 07071485 - 07071488, 07071501, 07088741 - 07088742, 07081124, 07088766, 07098769 - 07098773, 07098767 - 07098768, 07098774 - 07098775, 07108885 - 07108889, 07108883 - 07108884, 07108876 - 07108882, 07108890 - 07108891, 07101533 - 07101534, 07129027 - 07129028, RA000182, RA000184, 08029174 - 08029183, 08029187 - 08029188, 08029189 - 08029190, 09111694, 10011776, 10011777, 10011778, 10011779, 10011780, 10011781, 10011782, 10051792, 10051794, 10051796, 10051798, 10051800, 10051802, 10051804, 100610044, 100610045, 100610046, 100610047, 100610048, 100610049, 100610050, 100610051, 100610052, 100610053, 101010180, 101010182, 101010184, 101010186, 101010188, 101010190, 101010192, 101210252, BM005435 - BM005435, BM006121 - BM006121, BM008121 - BM008127
Distribution pattern: US States: MI
Derived failure modes
-
Unknown
The stopper block on the LFS bracket could be broken or missing. Without this stopper block, the flat screen attached to the LFS flat screen system can rotate beyond the original angle limit and it could cause excessive force to be given to the spring shaft of the LFS bracket. Due to the excessive force given to the spring shaft, the spring shaft could break and shear off.