Skip to content
Recall Observatory FDA recall evidence

Device product

Getinge Flow-i Anesthesia Systems Flow-i C20, Part no. 6677200 Flow-i C30, Part no. 6677300 Flow-i C40, Part no. 6677400

Z-0959-2023

December 22, 2022

Class I

Product summary

Firm
Getinge Usa Sales Inc
Event
Event 91339
Status
Ongoing
Classification
Class I
Quantity
8,016 Total (7,766 OUS; 250 US)
Official record key
device-enforcement:Z-0959-2023

Official wording

Reason: Due to a software bug, under certain conditions, pressure cannot be built up resulting in no ventilation. If gas delivery is stopped, a sustained decrease in delivered O2 concentration may lead to hypoventilation and hypoxia.

Code information: Part # UDI 6677200 7325710001349 Serial Numbers 1202-8027 Part # UDI 6677300 7325710001363 Serial Numbers 1058-1201 Part # UDI 6677400 7325710001318 Serial Numbers 2034-8446 Software versions 4.8.0, 4.8.1, 4.8.2, and 4.8.3.

Distribution pattern: US nationwide and Worldwide Distribution: Domestic distribution to AL, AR, AZ, CA, CO, DC, FL, GA, IN, MA, MI, MO, NY, OH, OR, PA, TX, VA, WA, WV. Foreign distribution worldwide.

Derived failure modes

  • Device software or design

    reason.device_software_design · v1.0.0

    software bug