MED-810A Zemits NDPrime Laser products

Z-0095-2024

October 09, 2023

Class II

Product summary

Firm
Advance-Esthetic LLC
Event
Event 93194
Status
Ongoing
Classification
Class II
Quantity
19
Official record key
device-enforcement:Z-0095-2024

Official wording

Reason: Advance-Esthetic LLC failed to comply with the applicable standards regarding Electronic Product Radiation Control (EPRC) regulations associated with the laser product performance standards and certification. This Recall provides correction by the firm regarding Zemits NdPrime Yag Tattoo Removal Laser. The product requires update or is missing the following in part: remote interlock connector, safety interlocks, manual reset mechanism, beam attenuator, tests for determination of compliance, certification, class IV designation and warning, labels, and labeling.

Code information: Model MED-810A Zemits NDPrime Laser products

Distribution pattern: US Nationwide Distribution

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Advance-Esthetic LLC failed to comply with the applicable standards regarding Electronic Product Radiation Control (EPRC) regulations associated with the laser product performance standards and certification. This Recall provides correction by the firm regarding Zemits NdPrime Yag Tattoo Removal Laser. The product requires update or is missing the following in part: remote interlock connector, safety interlocks, manual reset mechanism, beam attenuator, tests for determination of compliance, certification, class IV designation and warning, labels, and labeling.