Device product
RX Daytona Plus (with ISE)-IVD Clinical Chemistry Analyser Catalog Number: RX4040
Z-0198-2024
Product summary
- Event
- Event 93102
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 5 units
- Official record key
device-enforcement:Z-0198-2024
Official wording
Reason: RX Parameter rerun settings (Sample Volume Low/High), and Dilutions used for Auto reruns, not meeting sample re-run conditions when samples are above or below assay range and may result i issues with the re-run feature conditions when incorrectly. Potential to report incorrectly.
Code information: GTIN: 05055273207774 Lot/ Batch Number (1) Lot/ Batch Number 7241-0551 7241-0545 7241-0531 7241-0317 7241-0219
Distribution pattern: US Nationwide distribution including Puerto Rico
Derived failure modes
-
Unknown