Skip to content
Recall Observatory FDA recall evidence

Device product

stryker Tornier Perform Reversed Augmented Glenoid Lateralized Baseplate, 25mm, Offset +3mm, REF DWJ502

Z-0113-2024

September 15, 2023

Class II

Product summary

Firm
Tornier, Inc
Event
Event 93051
Status
Ongoing
Classification
Class II
Quantity
30 units
Official record key
device-enforcement:Z-0113-2024

Official wording

Reason: One lot of Tournier Perform Reversed Lateralized Baseplates were found to have been manufactured with 6mm spacers instead of 3mm spacers. This nonconformance prevents the central screw from being able to thread into the baseplate.

Code information: UDI/DI 10846832062017, Lot Number 1756123

Distribution pattern: US: CA, CO, IL, MA, MN, NC, OR, RI, SC, TN

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0