Skip to content
Recall Observatory FDA recall evidence

Device product

Ocean Water Seal Chest Drains (Ocean): Products REF Number Product Name (1)2002-000 DRAIN, OCEAN SINGLE W/AC,S; (2)2002-040 DRAIN, OCEAN SINGLE,PEDI CONNECTOR; (3)2002-100 DRAIN, OCEAN SINGLE W/AC; (4)2002-300 DRAIN, OCEAN SINGLE W/S; (5) 2002-400 DRAIN, OCEAN SINGLE ; (6)2012-320 DRAIN, OCEAN PEDI W/S; (7)2050-000 DRAIN, OCEAN BRU W/AC,S

Z-0316-2024

September 18, 2023

Class II

Product summary

Firm
Atrium Medical Corporation
Event
Event 93149
Status
Ongoing
Classification
Class II
Quantity
31270 cases: 29942 cases US; 1328 cases OUS
Official record key
device-enforcement:Z-0316-2024

Official wording

Reason: Instructions for Use (IFU) for the Atrium Ocean, Oasis, and Express chest drains do not provide sufficient precaution instruction for proper set up of catheter(s) and patient tube connections with single collection chamber chest drains

Code information: 1)2002-000 DRAIN, OCEAN SINGLE W/AC,S 20650862100017 (2)2002-040 DRAIN, OCEAN SINGLE,PEDI CONNECTOR 20650862100345 (3)2002-100 DRAIN, OCEAN SINGLE W/AC 20650862100093 (4)2002-300 DRAIN, OCEAN SINGLE W/S 20650862100109 (5) 2002-400 DRAIN, OCEAN SINGLE 20650862100215 (6)2012-320 DRAIN, OCEAN PEDI W/S 20650862101021 (7)2050-000 DRAIN, OCEAN BRU W/AC,S 20650862103018 All lots within labeled product expiry Manufacturing Dates: Any product manufactured on and after July 21, 2020 Distribution Dates: Any product shipped on and after August 6, 2020

Distribution pattern: Worldwide distribution - US Nationwide and the countries of Algeria, Australia, Austria, Bahrain, Belgium, Brazil, Brunei Darussalam, Canada, Chile, China, Czech Republic Ecuador, Egypt, Finland, France, Germany, Hong Kong, Iceland, Indonesia, Ireland, Israel, Italy, Jordan, Korea, Republic of Kuwait, Latvia, Lithuania, Luxembourg, Malaysia, Malta, Mauritius, Mexico, Montenegro, Netherlands, New Zealand, Norway, Pakistan, Peru, Poland, Portugal, Puerto Rico, Qatar, Russian Federation, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan Province of China, Thailand Tunisia, Ukraine, United Arab Emirates, United Kingdom, Uruguay.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Instructions for Use (IFU) for the Atrium Ocean, Oasis, and Express chest drains do not provide sufficient precaution instruction for proper set up of catheter(s) and patient tube connections with single collection chamber chest drains