Device product
Constellation Vision System, REF: 8065751150
Z-0099-2024
Product summary
- Event
- Event 93010
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 3
- Official record key
device-enforcement:Z-0099-2024
Official wording
Reason: Ophthalmic microsurgical vision systems have a new Printed Circuit Board (PCB) design that could lead to higher voltage to a PCB component, that could cause component failure and loss of pressure control for the use of cutters, scissors, and forceps, which could lead to an increased risk of microbial infection, tissue damage and progressive visual impairment.
Code information: UDI-DI: 00380657511501, Serial Numbers: 1403028701X, 1202865201X, 0802983301X
Distribution pattern: US: GA, IL, PR
Derived failure modes
-
Unknown
Ophthalmic microsurgical vision systems have a new Printed Circuit Board (PCB) design that could lead to higher voltage to a PCB component, that could cause component failure and loss of pressure control for the use of cutters, scissors, and forceps, which could lead to an increased risk of microbial infection, tissue damage and progressive visual impairment.