Iradimed Corporation MRidium MRI 1000 Series Infusion Sets, MRI Infusuui Set - Syringe Adapter Set, REF 1057-50,

Z-1310-2023

February 24, 2023

Class II

Product summary

Firm
Iradimed Corporation
Event
Event 91846
Status
Ongoing
Classification
Class II
Quantity
3285 boxes of 50 devices
Official record key
device-enforcement:Z-1310-2023

Official wording

Reason: Certain Iradimed 1057 Syringe Adapter Sets, used with the MRidium MRI infusion pump, have a problem relating to a defective injection molded part which can limit the effectiveness of the syringe venting function which then can result in occlusion.

Code information: UDI di 10856038001048, Lot Numbers: 21K50N, exp. 11/30/2023; 22A03N, exp. 01/31/2024; 22A59N, exp. 2/29/2024; 22B54N, exp. 2/29/2024; 22D15N, exp. 4/30/2024; 22D28N, exp. 4/30/2024; 22D58N, exp. 4/30/2024; 22I17N, exp. 9/30/2024; 031851, exp. 9/30/2024; 22J46N, exp. 10/31/2024; 032198, exp. 11/30/2024; 22L02N,exp. 12/31/2024; 032822, exp. 1/31/2025

Distribution pattern: Worldwide distribution - US Nationwide and the countries of Austria, Canada, Chile, Denmark, Germany, Great Britain, Hong Kong, Indonesia, Italy, Korea, Kuwait, Malaysia, Netherlands, Nicaragua, Oman, Pakistan, Peru, Philippines, Poland, Portugal, Qatar, Romania, Serbia, Singapore, Singapore, South Africa, Spain, Switzerland, Taiwan, Thailand, United Arab Emirates.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Certain Iradimed 1057 Syringe Adapter Sets, used with the MRidium MRI infusion pump, have a problem relating to a defective injection molded part which can limit the effectiveness of the syringe venting function which then can result in occlusion.