Skip to content
Recall Observatory FDA recall evidence

Device product

HIGH PRESSURE ROTATOR WITH MALE LUER LOCK, List Number MX496HP

Z-0743-2024

November 10, 2023

Class II

Product summary

Firm
Smiths Medical Asd Inc
Event
Event 93506
Status
Ongoing
Classification
Class II
Quantity
31,685 total
Official record key
device-enforcement:Z-0743-2024

Official wording

Reason: There is a manufacturing defect affecting specific rotator lots. Within this population, the inner diameter of the devices O-ring may be oversized affecting seal integrity. An inadequate seal may lead to a leak during infusion of medication which may potentially lead to under infusion of medication.

Code information: Lot Number: 4332915 4408772

Distribution pattern: Worldwide distribution - US Nationwide and the countries of Canada, UAE.

Derived failure modes

  • Labeling or packaging

    reason.labeling_packaging · v1.0.0

    inadequate seal
  • Manufacturing or process control

    reason.manufacturing_process_control · v1.0.0

    manufacturing defect