Skip to content
Recall Observatory FDA recall evidence

Device product

Ventlab, LLC SafeSpot manual resuscitators with integrated manometer, multiple accessory configurations to include: 1. SafeSpot Infant Resuscitator REF #s: SS3200OBPC-T, SS3204MB-D12, SS3200MB-2, SS3200MB-D1, SS3200OBP, SS3200MB-MMC, SS3200MBP-M00, SS3200MB-PW, SS3200MBP-2I, SS3200MBF-2, SS3204MBF-2, SS3200MB- PWS10, SS3206MCP-S1, SS3204MB, SS3204MC-P2, SS3200MB, SS3200MBP-2, SS3200MBF, SS3200MB-I, SS3200MBP, SS3200OB, SS3200MC, SS3200MC-P2; 2. SafeSpot Safe Spot Bag, D1 Mask with Inflation Bag Reservoir, Manometer, PEEP Valve, and 40cm Pop-Off REF # PRO-1927.

Z-0365-2024

October 11, 2023

Class II

Product summary

Firm
SunMed Holdings, LLC
Event
Event 93237
Status
Ongoing
Classification
Class II
Quantity
N/A
Official record key
device-enforcement:Z-0365-2024

Official wording

Reason: A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.

Code information: UDI-DI: 10889483104107, (Case UDI: 30889483104105, Item # PRO-1927); UDI-DI: 10889483102954, (Case UDI: 30889483102958, Item # SS3200MB); UDI-DI: 10889483092590, (Case UDI: 30889483092594, Item # SS3200MB-2); UDI-DI: 10889483104330, (Case UDI: 30889483104334, Item #SS3200MB-D1); UDI-DI: 10889483172322, (Case UDI: 30889483172326, Item # SS3200MBF); UDI-DI: 10889483104354, (Case UDI: 30889483104358, Item # SS3200MBF-2); UDI-DI: 10889483104545, (Case UDI: 30889483104549, Item # SS3200MB-I); UDI-DI: 10889483105771, (Case UDI: 30889483105775, Item # SS3200MB-MMC); UDI-DI: 10889483086919, (Case UDI: 30889483086913, Item # SS3200MBP); UDI-DI: 10889483104903, (Case UDI: 30889483104907, Item # SS3200MBP-2); UDI-DI: 10889483107324, (Case UDI: 30889483107328, Item # SS3200MBP-2I); UDI-DI: 10889483103777, (Case UDI: 30889483103771, Item # SS3200MBP-M00); UDI-DI: 10889483103111, (Case UDI: 30889483103115, Item # SS3200MB-PW); UDI-DI: 10889483086902, (Case UDI: 30889483086906, Item # SS3200MB- PWS10); UDI-DI: 10889483103784, (Case UDI: 30889483103788, Item # SS3200MC); UDI-DI: 10889483103128, (Case UDI: 30889483103122, Item # SS3200MC-P2); UDI-DI: 10889483107904, (Case UDI: 30889483107908, Item # SS3200OB); UDI-DI: 10889483086926, (Case UDI: 30889483086920, Item # SS3200OBP); UDI-DI: 10889483231142, (Case UDI: 30889483231146, Item # SS3200OBPC-T); UDI-DI: 10889483107331, (Case UDI: 30889483107335, Item # SS3204MB); UDI-DI: 10889483158913, (Case UDI: 30889483158917, Item # SS3204MB-D12); UDI-DI: 10889483104354, (Case UDI: 30889483104358, Item # SS3204MBF-2); UDI-DI: 10889483160497, (Case UDI: 30889483160491, Item # SS3204MC-P2); UDI-DI: 10889483111734, (Case UDI: 30889483111738, Item # SS3206MCP-S1. All lots manufactured from 9/21/2018 to present. Latest expiration Date 10/11/2028.

Distribution pattern: US Nationwide distribution including in the states of AR, AZ, CA¿CO, CT, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MN, NC, ND, NH, NJ, NM, NV, NY, OH, SC, TN, TX, UT, VA, & WV.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.