Skip to content
Recall Observatory FDA recall evidence

Device product

SOLTIVE SuperPulsed Laser Fibers Model Numbers TFL-FBX150S, TFL-FBX200S, TFL-FBX365S, TFL-FBX550S, TFL-FBX940S, TFL-FBX150BS, TFL-FBX200BS

Z-0748-2024

December 01, 2023

Class II

Product summary

Firm
Olympus Corporation of the Americas
Event
Event 93649
Status
Ongoing
Classification
Class II
Quantity
82,200 total units (5 per box)
Official record key
device-enforcement:Z-0748-2024

Official wording

Reason: Olympus identified inconsistencies in the Instructions for Use regarding cleaning the fiber tip and fiber cleaving and stripping instructions. Following the current instructions presents the potential for the fiber to break during use, which may cause burns, procedure delays, prolonged surgery, and foreign body in patient.

Code information: Models UDI/GTIN TFL-FBX150S 00821925043916 TFL-FBX200S 00821925043930 TFL-FBX365S 00821925043992 TFL-FBX550S 00821925044036 TFL-FBX940S 00821925044074 TFL-FBX150BS 00821925043879 TFL-FBX200BS 00821925043978 All lots affected

Distribution pattern: Worldwide distribution - US Nationwide and the countries of Canada, Europe, Hong Kong, Australia, Japan, Singapore, and Latin America.

Derived failure modes

  • Foreign material or chemical contamination

    reason.foreign_material_chemical_contamination · v1.0.0

    foreign body