Skip to content
Recall Observatory FDA recall evidence

Device product

Naturalyte Liquid Acid concentrate bottles, Product Code 08-1251-1 - Product Usage: used in conjunction with Fresenius Renal Technologies Sodium Bicarbonate Concentrate in a compatible 45X dilution three-stream artificial kidney (hemodialysis) machine.

Z-0831-2021

November 23, 2020

Class II

Product summary

Firm
Fresenius Medical Care Holdings, Inc.
Event
Event 86939
Status
Ongoing
Classification
Class II
Quantity
1686 cases
Official record key
device-enforcement:Z-0831-2021

Official wording

Reason: The deviation in conductivity for sodium level exceeds 0.6mS/cm from the theoretical conductivity (TCD) or approximately 6mEq/L of sodium from prescriptions that was previously communicated. Calcium concentration is below manufacturing specification.

Code information: Lot Numbers: 20LXAC063 20NXAC008 Additional Affected lots as of 3/17/21: 20NXAC005 20LXAC098 20NXAC039 20NXAC008 20LXAC063

Distribution pattern: US Nationwide distribution.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0