Device product
Medtronic HVAD Monitor Data Cable, Product Number: 1575
Z-1339-2021
Product summary
- Firm
- Heartware, Inc.
- Event
- Event 87446
- Status
- Ongoing
- Classification
- Class I
- Quantity
- 1337 units (1110 US)
- Official record key
device-enforcement:Z-1339-2021
Official wording
Reason: There is the potential for the Controller power cables (AC Adapter, DC Adapter, and Battery cables), HVAD Alarm Adapter, and HVAD Monitor data cables to cause damage to the HVAD Controller ports when inserting the cables into the Controller ports, due to wear from misalignment of the cables to the port.
Code information: All lot codes
Distribution pattern: Worldwide
Derived failure modes
-
Unknown
There is the potential for the Controller power cables (AC Adapter, DC Adapter, and Battery cables), HVAD Alarm Adapter, and HVAD Monitor data cables to cause damage to the HVAD Controller ports when inserting the cables into the Controller ports, due to wear from misalignment of the cables to the port.