Device product
Amicus MNC Apheresis Kit - Double Needle - Product Usage: intended for use in the collection of blood components and mononuclear cells.
Z-1343-2021
Product summary
- Firm
- Fenwal Inc
- Event
- Event 87407
- Status
- Terminated
- Classification
- Class II
- Quantity
- 4488 units
- Official record key
device-enforcement:Z-1343-2021
Official wording
Reason: Some lots of Amicus MNC Apheresis and Amicus Exchange kits have leaking centrifuge packs during a procedure on the Amicus Separator using therapeutic protocols only. The defect is a blood leak at the boot, elbow, or in the channel on the separation chamber of the centrifuge pack.
Code information: Product Code X6R2326; UDI: 04086000100144; Batch Numbers FA20E25234 ((Exp. 05/31/2022), FA20F23120 (Exp. 06/30/2022), FA20I21268 (Exp. 09/30/2022), and FA20J19246 (Exp. 10/31/2022).
Distribution pattern: US Nationwide distribution in the states of GA, TX, TN, MD, CA, KS, MN, MA, NM, NC, LA, CO, KY, NV, PA, UT, MI, NJ, IL, MD, OH, NY, FL.
Derived failure modes
-
Unknown
Some lots of Amicus MNC Apheresis and Amicus Exchange kits have leaking centrifuge packs during a procedure on the Amicus Separator using therapeutic protocols only. The defect is a blood leak at the boot, elbow, or in the channel on the separation chamber of the centrifuge pack.