Device product

DOLOMITE GLOSS ROLLATOR - intended to support during walking and may also serve as a resting seat. Model Number(s): 1654807 - DOLOMITE GLOSS 680 ROLLATOR 1655081 - DOLOMITE GLOSS 600 ROLLATOR 1655082 - DOLOMITE GLOSS 520 ROLLATOR 1655083 - DOLOMITE GLOSS 450 ROLLATOR

Z-2445-2023

June 15, 2023

Class II

Product summary

Firm
Invacare Corporation
Event
Event 92645
Status
Ongoing
Classification
Class II
Quantity
586 devices
Official record key
device-enforcement:Z-2445-2023

Official wording

Reason: Premature failure of the seat during use, the plastic eyelets of the folding mechanism of the device's seat can break, and in a worst-case crossbar also breaks and the Rollator can collapse causing injury to user

Code information: Serial number in the range: 20BHL0002 to 21IHL1910

Distribution pattern: US Nationwide distribution.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Premature failure of the seat during use, the plastic eyelets of the folding mechanism of the device's seat can break, and in a worst-case crossbar also breaks and the Rollator can collapse causing injury to user